Radiography: Automatic Exposure Control (AEC)

160 flashcardsby Z
Study these flashcardsFree · spaced repetition · iPhone, Android, Mac & web

All cards

1
What does automatic exposure control (AEC) determine during a radiographic exposure?
When to terminate the exposure
2
What component detects radiation in a typical AEC system?
Ionization chamber
3
In an AEC exposure, where must the examined anatomy be positioned?
Directly over the selected ionization chamber
4
How many ionization chambers are typically present in an AEC system?
3
5
Where is the center AEC chamber located relative to the other two?
Centered on the IR, with the left and right chambers slightly higher
6
Why is the central AEC chamber positioned below the duodenum and transverse colon for many abdominal studies?
To reduce interference from gastric and bowel gas
7
With three AEC chambers, how many nonempty chamber-selection combinations are possible?
7
8
What can a collimator light insert identify for an AEC system?
1. Chamber size
2. Chamber shape
3. Chamber location
9
Which AEC chamber tends to dominate when multiple chambers are selected?
The chamber receiving the most radiation
10
What AEC setting changes receptor exposure before acquisition rather than modifying the displayed image afterward?
Density control
11
Which AEC density-control settings generally indicate normal receptor exposure?
0 or N
12
What is the difference between major and minor AEC density controls?
Major makes large exposure adjustments, whereas minor fine-tunes exposure
13
How does AEC density control differ from digital post-processing?
Density control changes radiation exposure, whereas post-processing changes image appearance
14
Why should AEC density control not be used to compensate for poor positioning?
It cannot correct which anatomy covers the active chamber
15
Which four selections does the radiographer still make when using an AEC technique chart?
1. kVp; 2. mA; 3. SID; 4. Chamber selection
16
Under AEC, which parameter is determined by the system rather than manually preset?
A. kVp
B. mA
C. Exposure time
D. SID
C. Exposure time
17
During a table examination, the upright Bucky AEC chambers are accidentally selected. Why can exposure be unreliable?
The active chambers are not behind the imaged anatomy
18
A chest examination has fluid-filled lungs that attenuate more radiation over the AEC chamber. How does AEC respond?
It increases exposure time
19
An emphysema patient allows more radiation to reach an AEC chamber. What happens to exposure time?
AEC terminates the exposure sooner
20
How can excessively tight collimation cause AEC overexposure?
It cuts off the chamber so AEC continues exposing too long because it thinks the anatomy is denser than usual
21
How can overly wide collimation cause premature AEC termination?
Additional scatter reaches the chamber
22
A very thin anatomical part requires less exposure time than the AEC system can deliver. What adjustment is recommended?
Decrease mA
23
What is the approximate minimum response time of modern AEC systems?
0.001 second
24
Which system combines computerized, exam-specific technique presets with automatic exposure control?
Anatomically Programmed Radiography (APR)
25
What is anatomically programmed radiography (APR)?
A computerized system selecting technique factors based on the examination, combined with AEC
26
If no ionization chamber is sufficiently covered by anatomy, which approach is preferable?
A. Manual exposure technique
B. All three AEC chambers
C. Maximum AEC density setting
D. Widest possible collimation
A. Manual exposure technique
27
When an unused AEC chamber is not covered by anatomy but another selected chamber is covered, what can the radiographer do?
Deactivate the uncovered chamber
28
True or False: AEC eliminates the need for a radiographic technique chart.
False. AEC still requires a technique chart
29
An AEC chamber is beneath mostly unattenuated lung rather than the intended dense anatomy. What is a likely result?
AEC stops early, underexposing the dense anatomy
30
Which AEC mistake is most directly associated with incorrect chamber selection?
A. Wrong receptor exposure
B. Improved spatial resolution
C. Shorter SID
D. Reduced magnification
A. Wrong receptor exposure
31
True or False: AEC automatically determines what anatomical structure is positioned over an active chamber.
False. It only responds to radiation reaching the chamber
32
Which aspect of exposure does a density control adjust, and when does the adjustment occur?
Radiation reaching the IR; before the image is acquired
33
A radiographer needs less exposure from AEC without changing positioning. Which density-control direction is appropriate?
Toward the negative setting
34
An AEC chamber is beneath unattenuated lung instead of the intended denser structure. What is the likely result?
Early termination and underexposure of the intended structure
35
What primarily determines whether AEC will produce consistent diagnostic images?
Proper positioning and chamber selection
36
Who developed one of the first ionization chamber systems in 1942?
Russell H. Morgan in 1942
37
What is the typical thickness of the AEC ionization chambers?
5 mm
38
What limitation applies to collimator light inserts that indicate AEC chamber positions?
They are accurate only at specific SIDs
39
Why can overreliance on AEC weaken a radiographer's technical skills?
AEC does not replace knowledge of manual exposure techniques or positioning
40
What happens to AEC exposure time when a positive density-control setting is selected, with other factors unchanged?
It increases
41
Which adjustment changes patient radiation dose rather than simply changing the appearance of an acquired image?
A. Window level
B. Window width
C. AEC density control
D. Post-processing
C. AEC density control
42
Why might AEC produce excessive exposure in unaffected areas of a chest image when fluid overlies the active chamber?
The fluid attenuates radiation, prolonging exposure
43
A thin part requires an exposure shorter than the AEC minimum response time. What exposure error may occur if mA is not reduced?
Overexposure
44
Which factors increase Compton scatter production?
Higher kVp; larger field size; greater patient thickness
45
What happens to scatter production when x-ray field size increases?
Increases
46
What happens to scattered radiation when the beam is collimated to a smaller field?
Decreases
47
What happens to scatter production as kVp increases?
Increases
48
How does increased scatter generally affect radiographic image contrast?
Decreases contrast
49
Which type of contrast has few shades of gray and a short grayscale?
High or short-scale contrast
50
Which window width corresponds to higher displayed image contrast?
Narrow window width
51
Why can anatomical compression reduce scatter during imaging?
It reduces tissue thickness
52
What does tissue compression do to object-to-image distance (OID)?
Decreases OID
53
How does increasing the atomic number of the irradiated material affect scatter?
Decreases scatter as photoelectric absorption becomes more likely
54
What two common radiopaque contrast substances are used in radiography?
Iodine; barium sulfate
55
What does subtraction imaging remove to emphasize injected contrast material?
Background anatomy
56
What is the most common beam-restricting device used in general radiography?
Collimator
57
Name the three basic beam-restricting devices.
1. Aperture diaphragm
2. Cone/cylinder
3. Collimator
58
A flat lead sheet with a central opening attached to the x-ray tube describes which beam restrictor?
Aperture diaphragm
59
What is the principal drawback of an aperture diaphragm?
Increased penumbra (geometric unsharpness)
60
What does penumbra mean in radiographic imaging?
Geometric unsharpness along an image edge
61
How can penumbra associated with an aperture diaphragm be reduced?
Use a longer SID
62
What is off-focus (extrafocal or stem) radiation?
Radiation produced away from the intended focal spot
63
A circular aperture diaphragm with an extended metal tube describes which beam restrictor?
Cone or cylinder
64
Which beam restrictor is more effective at reducing penumbra?
A. Aperture diaphragm
B. Flared cone
C. Cylinder
D. Lead mask
C. Cylinder
65
What characteristic allows a collimator to adapt to many different examinations?
Adjustable field size
66
What material forms the shutters of a radiographic collimator?
Lead
67
Which collimator shutters primarily reduce off-focus radiation?
Upper shutters
68
Which collimator shutters primarily reduce penumbra?
Bottom shutters
69
What does the collimator light-localizing system use to project the approximate x-ray field?
Light source and mirror
70
What does PBL stand for in beam restriction?
Positive Beam Limitation
71
What is another term for a positive beam limitation device?
Automatic collimator
72
What happens when a cassette or imaging plate enters a Bucky equipped with automatic collimation?
The light field is automatically adjusted to the receptor size
73
Why should the light field not extend beyond the image receptor?
Unnecessary patient irradiation and image degradation
74
What type of filtration does the collimator contribute?
A. Inherent
B. Compensating
C. Total only
D. Added
D. Added
75
Which two ancillary beam-restricting devices are discussed?
Lead blocker; lead mask
76
What is a lead blocker typically made of?
Lead-impregnated rubber sheet
77
Where is a lead mask typically attached for a specific field shape?
End of the collimator
78
Why may a lead blocker help with a lateral lumbar-spine radiograph?
It absorbs scatter originating in posterior soft tissue
79
Which beam-limiting device combines adjustable lead shutters and a projected light field?
A. Collimator
B. Aperture diaphragm
C. Fixed cone
D. Lead mask
A. Collimator
80
True or False: Opening the collimator field wider generally reduces scatter reaching the receptor.
False. A wider field generally increases scatter
81
Which two physical factors determine the volume of irradiated tissue?
Field size; patient thickness
82
Which tissue produces more scatter: bone or soft tissue?
Soft tissue
83
In which imaging modality is anatomical compression especially common?
Mammography
84
What is a major patient-dose benefit of restricting the x-ray beam to the anatomy of interest?
Less tissue is irradiated
85
Which three advantages are associated with aperture diaphragms?
Simple design; low cost; ease of use
86
For which applications are aperture diaphragms particularly suitable?
Chest radiography; fixed trauma units; dental equipment
87
What accessory permits adjustment of an otherwise fixed cone field size?
Extension sleeve

*can allow radiation field size to be adjusted
88
How does a lead blocker differ from a lead mask in typical placement?
A blocker is placed on the table, whereas a mask is secured to the collimator
89
Which patient group benefits especially from lead blockers because of increased scatter production?
Larger patients
90
Which two photographic properties determine visibility of detail in the radiographic image?
1. Density (brightness)
2. Contrast
91
What word describes image density when a digital radiograph is viewed on a monitor?
Brightness
92
Which digital post-processing controls change displayed brightness and grayscale?
Window level
93
For radiographic images, what is the main controlling factor of image receptor exposure?
mAs
94
What is the approximate minimum change in mAs that produces a perceptible change in film density?
30%
95
Why is doubling or halving mAs preferred over a tiny adjustment when repeating an exposure?
Making a bold change to sufficiently change the receptor exposure is a good idea to prevent more unnecessary exposures
96
What is the step-in-time rule, and why should it be avoided?
Changing mAs by an arbitrary small time step; equipment differences make results unreliable
97
If you want to change IR exposure without changing contrast, which technical factor should you primarily alter?
A. kVp
B. SID
C. mAs
D. Grid ratio
C. mAs
98
What does raising kVp generally do to receptor exposure, with other factors constant?
Increases it
99
What does a 15% increase in kVp do to image receptor exposure according to the 15% rule?
Approximately doubles it
100
What does a 15% decrease in kVp do to image receptor exposure according to the 15% rule?
Approximately halves it
101
Why is adjusting kVp to change exposure different from adjusting mAs?
Changing kVp also changes radiographic contrast
102
An elbow was exposed at 60 kVp. Using the 15% rule, what whole-number kVp approximately halves IR exposure?
51 kVp
103
An abdomen was exposed at 75 kVp. Using the 15% rule, what whole-number kVp approximately doubles IR exposure?
86 kVp
104
A distal femur was exposed at 65 kVp. Using the 15% rule, what whole-number kVp approximately halves IR exposure?
55 kVp
105
Compared with a single-phase generator at the same selected factors, what happens to receptor exposure with a high-frequency generator?
It increases
106
When converting from single-phase to high-frequency generation, which mAs conversion factor maintains exposure?
0.5
107
When converting from high-frequency to single-phase generation, which mAs conversion factor maintains exposure?
2.0
108
A satisfactory single-phase exposure uses 20 mAs. What mAs maintains exposure with a high-frequency generator?
10 mAs
109
A satisfactory high-frequency exposure uses 4 mAs. What mAs maintains exposure with a single-phase generator?
8 mAs
110
What happens to film density when focal-spot size is changed on a properly calibrated radiographic unit?
Negligible change
111
High tube current causes an unwanted increase in effective focal-spot size from reduced electron focusing. What is this called?
Blooming
112
Where is beam intensity greater because of the anode heel effect?
Cathode end
113
For an object thicker at one end, where should the thicker portion be positioned to use the anode heel effect?
Toward the cathode end
114
When is the anode heel effect more pronounced: short SID or long SID?
Short SID
115
What happens to radiation intensity at the IR when distance increases with technique factors unchanged?
Decreases according to the inverse-square law
116
Which relationship calculates how exposure intensity changes with distance?
I₁/I₂ = D₂²/D₁²

inverse square law
117
An exposure measures 180 mR at 40 inches. What intensity is expected at 72 inches with the same technique?
55.6 mR
118
What formula adjusts mAs to maintain receptor exposure after changing SID?
New mAs = Original mAs × (new SID/old SID)²

Direct square law
Exposure maintence formula
119
A satisfactory exposure uses 20 mAs at 72 inches. What mAs maintains exposure at 40 inches?
6.17 mAs
120
A satisfactory exposure uses 10 mAs at 40 inches. What mAs maintains exposure at 52 inches?
16.9 mAs
121
What happens to receptor exposure when filtration is increased without changing other technique factors?
Decreases
122
What minimum total filtration is recommended above 70 kVp?
2.5 mm aluminum
123
When beam restriction is increased through tighter collimation, what happens to scatter and IR exposure?
Both decrease
124
As anatomical part thickness increases, what happens to IR exposure with the same technique?
Decreases
125
How should a body part inside a cast be treated when planning exposure technique?
As a thicker or denser anatomical part
126
How does adding a grid affect IR exposure at unchanged factors?
It decreases because the grid absorbs radiation
127
When using a grid with a higher conversion factor, what must happen to mAs to maintain receptor exposure?
Increase mAs
128
Which formula maintains exposure when changing grids?
New mAs = Old mAs × (new GCF/old GCF)
129
As film-screen relative speed increases, what happens to mAs needed for equal density?
Decreases
130
What is the relative-speed exposure maintenance formula?
New mAs = Old mAs × (old relative speed/new relative speed)
131
An acceptable image uses 10 mAs with a 400-speed system. What mAs is required with a 200-speed system?
20 mAs
132
Why can digital post-processing fail to rescue a severely underexposed image?
You can’t retrieve image information by changing the appearance of the image if the image information was never there to begin with
133
Which pathological change generally increases IR exposure with unchanged factors?
A. Additive pathology
B. Destructive pathology
C. Increased patient thickness
D. Higher grid ratio
B. Destructive pathology
134
Which pathological type decreases IR exposure: additive or destructive?
Additive pathology
135
How does radiolucent contrast media affect IR exposure?
Increases it
136
How does radiopaque contrast media affect IR exposure?
Decreases it
137
What effect does increasing generator waveform pulses have on IR exposure?
Increases it
138
Which imaging modality uses the term density for the degree of image blackening?
Analog screen-film radiography
139
What distinguishes a soft-copy radiograph from a hard-copy radiograph?
Soft copy is displayed digitally, whereas hard copy is printed on film
140
Approximately how many shades of gray can a person distinguish at one time?
30
141
What is the purpose of a preset, examination-specific digital image histogram?
To guide image processing for that examination
142
Why can acceptable digital image brightness be misleading when evaluating an exposure?
Post-processing can alter brightness without restoring missing information
143
Can a radiographic image have exposure without contrast, or contrast without exposure?
Exposure without contrast is possible, but contrast requires exposure
144
What is the relationship of kVp to IR exposure: controlling or influencing factor?
Influencing factor
145
If kVp is increased by 15% but receptor exposure must remain approximately unchanged, what should happen to mAs?
Halve mAs
146
Which generator produces lower average photon energy at comparable selected kVp?
A. High frequency
B. Three-phase 12-pulse
C. Single-phase
C. Single-phase
147
What should be done if focal-spot blooming causes noticeable density differences?
Perform quality control and evaluate the x-ray tube
148
By up to what percentage can exposure be greater at the cathode end than the anode end?
45%
149
How does opening the collimator field affect the prominence of the anode heel effect?
Makes it more pronounced
150
What is the other name for the exposure maintenance formula relating mAs and SID?
Direct square law
151
An exposure measures 240 mR at 72 inches. What intensity is expected at 40 inches with unchanged technical factors?
777.6 mR

*Indirect square law
152
An exposure measures 110 mR at 52 inches. What intensity is expected at 40 inches with unchanged technical factors?
185.9 mR
153
A satisfactory image uses 40 mAs at 72 inches. What mAs maintains receptor exposure at 45 inches?
15.63 mAs
154
What minimum total filtration is recommended at 50–70 kVp?
1.5 mm aluminum
155
What minimum total filtration is recommended below 50 kVp?
0.5 mm aluminum
156
Why is filtration beyond 3 mm aluminum considered a point of diminishing returns?
Further reduction in entrance skin exposure may not justify increased tube loading
157
How does increasing tissue atomic number affect receptor exposure, assuming comparable thickness and unchanged technique?
Decreases it
158
Which factor has a negligible overall effect on IR exposure?
A. Increasing filtration
B. Increasing grid ratio
C. Anode heel effect
D. Increasing distance
C. Anode heel effect
159
What happens to film density when developer time, temperature, or chemical activity increases?
Increases
160
What effect does contaminated film-processing solution have on film density?
Decreases it
Study all 160 cards for free

More Medicine & health flashcards

KNES 25947 cardsMedicine & health · EnglishZanderpredlaktie + ruka37 cardsMedicine & health · LatinBronislavaDigestive system30 cardsMedicine & health · EnglishKayleighExámen24 cardsMedicine & health · SpanisheveSLE and Sjogens Syndrome20 cardsMedicine & health · EnglishAveryHéterogénitées du menteau niegeux 15 cardsMedicine & health · FrenchLilipharmacodynamie10 cardsMedicine & health · FrenchClaraAnatomy Basics: Planes and Landmarks186 cardsMedicine & health · EnglishNjEl aparato digestivo (U.3 Parte 2)138 cardsMedicine & health · SpanishTomateChapter 21 — Fluid and Electrolyte Balance95 cardsMedicine & health · EnglishlaviniaPregnancy Hypertension: Medications and Care74 cardsMedicine & health · EnglishAlaa.m.Anatomy & Physiology: Digestive System57 cardsMedicine & health · EnglishIndiamara

All Medicine & health decks

Study these decks in the app

Free on iPhone, Android, Mac and the web. Add any deck to your account in one tap.

Download on the App StoreGet it on Google Play
Study all 160 cards for free