Radiography: Automatic Exposure Control (AEC) Chapter 7

43 flashcardsby Z
Study these flashcardsFree · spaced repetition · iPhone, Android, Mac & web

All cards

1
What does automatic exposure control (AEC) determine during a radiographic exposure?
When to terminate the exposure
2
What component detects radiation in a typical AEC system?
Ionization chamber
3
In an AEC exposure, where must the examined anatomy be positioned?
Directly over the selected ionization chamber
4
How many ionization chambers are typically present in an AEC system?
3
5
Where is the center AEC chamber located relative to the other two?
Centered on the IR, with the left and right chambers slightly higher
6
Why is the central AEC chamber positioned below the duodenum and transverse colon for many abdominal studies?
To reduce interference from gastric and bowel gas
7
With three AEC chambers, how many nonempty chamber-selection combinations are possible?
7
8
What can a collimator light insert identify for an AEC system?
1. Chamber size; 2. Chamber shape; 3. Chamber location
9
Which AEC chamber tends to dominate when multiple chambers are selected?
The chamber receiving the most radiation
10
What AEC setting changes receptor exposure before acquisition rather than modifying the displayed image afterward?
Density control
11
Which AEC density-control settings generally indicate normal receptor exposure?
0 or N
12
What is the difference between major and minor AEC density controls?
Major makes large exposure adjustments, whereas minor fine-tunes exposure
13
How does AEC density control differ from digital post-processing?
Density control changes radiation exposure, whereas post-processing changes image appearance
14
Why should AEC density control not be used to compensate for poor positioning?
It cannot correct which anatomy covers the active chamber
15
Which four selections does the radiographer still make when using an AEC technique chart?
1. kVp; 2. mA; 3. SID; 4. Chamber selection
16
Under AEC, which parameter is determined by the system rather than manually preset?
  • kVp
  • mA
  • Exposure time
  • SID
17
During a table examination, the upright Bucky AEC chambers are accidentally selected. Why can exposure be unreliable?
The active chambers are not behind the imaged anatomy
18
A chest examination has fluid-filled lungs that attenuate more radiation over the AEC chamber. How does AEC respond?
It increases exposure time
19
An emphysema patient allows more radiation to reach an AEC chamber. What happens to exposure time?
AEC terminates the exposure sooner
20
How can excessively tight collimation cause AEC overexposure?
It cuts off the chamber so AEC continues exposing too long
21
How can overly wide collimation cause premature AEC termination?
Additional scatter reaches the chamber
22
A very thin anatomical part requires less exposure time than the AEC system can deliver. What adjustment is recommended?
Decrease mA
23
What is the approximate minimum response time of modern AEC systems?
0.001 second
24
Which system combines computerized, exam-specific technique presets with automatic exposure control?
Anatomically Programmed Radiography (APR)
25
What is anatomically programmed radiography (APR)?
A computerized system selecting technique factors based on the examination, combined with AEC
26
If no ionization chamber is sufficiently covered by anatomy, which approach is preferable?
  • Manual exposure technique
  • All three AEC chambers
  • Maximum AEC density setting
  • Widest possible collimation
27
When an unused AEC chamber is not covered by anatomy but another selected chamber is covered, what can the radiographer do?
Deactivate the uncovered chamber
28
True or False: AEC eliminates the need for a radiographic technique chart.
False. AEC still requires a technique chart
29
An AEC chamber is beneath mostly unattenuated lung rather than the intended dense anatomy. What is a likely result?
AEC stops early, underexposing the dense anatomy
30
Which AEC mistake is most directly associated with incorrect chamber selection?
  • Wrong receptor exposure
  • Improved spatial resolution
  • Shorter SID
  • Reduced magnification
31
True or False: AEC automatically determines what anatomical structure is positioned over an active chamber.
False. It only responds to radiation reaching the chamber
32
Which aspect of exposure does a density control adjust, and when does the adjustment occur?
Radiation reaching the IR; before the image is acquired
33
A radiographer needs less exposure from AEC without changing positioning. Which density-control direction is appropriate?
Toward the negative setting
34
An AEC chamber is beneath unattenuated lung instead of the intended denser structure. What is the likely result?
Early termination and underexposure of the intended structure
35
What primarily determines whether AEC will produce consistent diagnostic images?
Proper positioning and chamber selection
36
Who developed one of the first ionization chamber systems in 1942?
Russell H. Morgan
37
What is the typical thickness of the AEC ionization chambers?
5 mm
38
What limitation applies to collimator light inserts that indicate AEC chamber positions?
They are accurate only at specific SIDs
39
Why can overreliance on AEC weaken a radiographer's technical skills?
AEC does not replace knowledge of manual exposure techniques or positioning
40
What happens to AEC exposure time when a positive density-control setting is selected, with other factors unchanged?
It increases
41
Which adjustment changes patient radiation dose rather than simply changing the appearance of an acquired image?
  • Window level
  • Window width
  • AEC density control
  • Post-processing
42
Why might AEC produce excessive exposure in unaffected areas of a chest image when fluid overlies the active chamber?
The fluid attenuates radiation, prolonging exposure
43
A thin part requires an exposure shorter than the AEC minimum response time. What exposure error may occur if mA is not reduced?
Overexposure
Study all 43 cards for free

More Medicine & health flashcards

Cardiotoxic Plants in Veterinary Medicine19 cardsMedicine & health · EnglishDaniela15. kémiai a és irradiációs karcinogenezis14 cardsMedicine & health · HungarianRitikeDiabetes mellitus9 cardsMedicine & health · HungarianRitikeNervous System203 cardsMedicine & health · EnglishEshaAbdomen130 cardsMedicine & health · EnglishEshaCSF91 cardsMedicine & health · EnglishCHRISTYRadiography: IR Exposure (Chapter 26)71 cardsMedicine & health · EnglishZPulmonary Pathology56 cardsMedicine & health · EnglishMadiruth2Hematopoietic and Lymphoproliferative Neoplasia46 cardsMedicine & health · EnglishDanielaMicosis subcutáneas36 cardsMedicine & health · SpanishMajoHUMERUS in STERNUM30 cardsMedicine & health · LatinemGrundlagen der Anatomie und Physiologie23 cardsMedicine & health · GermanGaby

All Medicine & health decks

Study these decks in the app

Free on iPhone, Android, Mac and the web. Add any deck to your account in one tap.

Download on the App StoreGet it on Google Play
Study all 43 cards for free