Exam 2 - Detection and Measurements

40 flashcardsby Lindsey
Study these flashcardsFree · spaced repetition · iPhone, Android, Mac & web

All cards

1
Types of radiation detectors
-gas-filled
-Scintillation
-Solid-state, Semiconductor
-others:
–Cloud chambers
– Bubble Chambers
– Superheated Drop
– Cryogenic
– AMANDA and IceCube
2
Detector Parameters
• Detection Mode
• Operation Mode
• Dead Time / Interaction Rate
• Paralyzable / Non-Paralyzable
• Efficiency
3
Describe the detection modes (how they work)
– Counters – detect number of interaction events
– Spectrometers – detect number of events as a
function of energy
– Dosimeters – detect accumulated energy by all
interactions
4
Describe the operation modes (how they tell us)
Pulse – detects (and generally counts) individual interactions
Current –individual interactions averaged to produce current
5
Describe dead time
Maximum rate at which data can be recorded
6
relationship between true count rate and recorded ramp rate
n = m/(1-(m*tau))

m: recorded ramp rate
n: true count rate
7
What determines dead time
Duty cycle of slowest component determines
dead time
8
What detectors have longest dead time
Geiger-Müller (GM) counting
• In multichannel analyzer systems - analog-to-digital converter
9
Typical GM dead time
10-100 ms
10
Most other sytems' dead time
<3-8 ms
11
Interaction rate in pulse mode and why its important
• In pulse mode, events must be separated by
more than the dead time to be detected
• A second interaction in this interval will not be
detected
• A second interaction very close to the first
interaction may distort the signal from the first
interaction
12
Paralyzable vs nonparalyzable
• Systems in which dead-time events extend
dead time are paralyzable. Otherwise,
systems are nonparalyzable.
• At very high interaction rates, paralyzable
systems will not detect any interactions
after the first, causing the detector to
indicate a count rate of zero
13
What operation mode do spectrometers often operate in
pulse mode
14
How Spectrometers work
• Amplitude of each pulse is proportional to the
energy deposited in the detector by the
interaction causing that pulse
• The energy deposited by an interaction is not
always the total energy of the incident particle or
photon
• Pulse height spectrum - graph of the number of
interactions depositing a particular amount of
energy as a function of energy
15
Current mode operation
In current mode, all information regarding
individual interactions is lost, but these systems
can be designed with no dead time
• If the electrical charge collected from each
interaction is proportional to the energy
deposited by that interaction, then the net
current is proportional to the dose rate in the
detector material
16
Which operation mode is used for detectors subjected to very high interaction rates
current mode operation
17
Detection efficiency
• The efficiency (sensitivity) of a detector is
a measure of its ability to detect radiation
• Efficiency of a detection system operated
in pulse mode is defined as the probability
that a particle or photon emitted by a
source will be detected
18
overall efficiency
#detected/#emitted AND geometric efficiency*intrinsic efficienct
19
geometric efficiency
#reaching detector/#emitted
20
intrinsic efficiency
#detected/#reaching detector

• Often called the quantum detection efficiency or
QDE
• Determined by the energy of the photons and
the atomic number, density, and thickness of the
detector
21
For a parallel beam of monoenergetic photons incident on a detector of uniform thickness what is the intrinsic efficienct equation
intrinsic efficiency = 1-e^(-mu*x)
22
Gas-filled detectors
• A gas-filled detector comprises gas between two
oppositely charged electrodes
• Ionizing radiation produces ion pairs in the gas
• Positive ions (cations) migrate to negative
electrode (cathode); electrons or anions migrate
to positive electrode (anode)
23
In most detectors, what is the anode and what is the cathode
anode is wire inside the container
cathode is the wall of the container that holds the gas
24
3 types of gas-filled detectors
– Ionization chambers
– Proportional counters
– Geiger-Müller (GM) counters
25
What primarily determines the difference between the 3 types of gas-filled detectors
The voltage applied between the 2 electrodes
26
benefit of ionization chambers
wider range of physical shape (parallel plates, concentric cylinders, etc.)
27
which gas filled detectors must have a thin wire anoed? Why?
Proportional counters and GM counters (look up answer in the book)
28
For ionization chambers, when is the amoung of current produced proportional to the exposure rate
When the gas is air and the walls of the chamber are a material whose effective atomic number is similar to air
29
Where are air-filled ion chambers used
-Portable survey meters
-performing QA testing of diagnostic and therapeutic x-ray machines
-detectors in most x-ray phototimers
30
Why do ionization chambers have low intrinsic efficiencies
Low density of gases and low atomic numbers of most gases
31
Tell me about proportional counters
•Must contain a gas with specific properties
• Commonly used in standards laboratories,
health physics laboratories, and for
physics research
• Seldom used in medical centers
32
GM counters
• GM counters also must contain gases with
specific properties
• Gas amplification produces billions of ion pairs
after an interaction – signal from detector
requires little amplification
• Often used for inexpensive survey meters
• In general, GM survey meters are inefficient
detectors of x-rays and gamma rays
• Over-response to low energy x-rays – partially
corrected by placing a thin layer of higher atomic
number material around the detector
• GM detectors suffer from extremely long dead
times – seldom used when accurate
measurements are required of count rates
greater than a few hundred counts per second
• Portable GM survey meter may become
paralyzed in a very high radiation field – should
always use ionization chamber instruments for
measuring such fields
33
Scintillation detectors
• Scintillators are used in conventional film-screen
radiography, many digital radiographic
receptors, fluoroscopy, scintillation cameras,
most CT scanners, and PET scanners
• Scintillation detectors consist of a scintillator and
a device, such as a PMT, that converts the light
into an electrical signal

• Amount of light emitted after an interaction
increases with energy deposited by the
interaction
• May be operated in pulse mode as
spectrometers
• High conversion efficiency produces
superior energy resolution
34
Inorganic Scintillation Mechanism
– Depends on crystalline structure and hence must be crystalline
or poly-crystalline
– Involve high Z elements (compared to organics), especially as
activator/dopants, and are therefore effective gamma absorbers
– Generally hydroscopic and fragile
35
Organic Scintillation Mechanism
– Depend on molecular rather than crystalline structure, and hence
more easily fabricated.
– Less sensitive generally and especially to gamma radiation
– Low Z-numbers produce less backscatter
– Good for electron and beta detection and for fast neutron
detection (fast neutrons produce fast H and C that ionize as they
slow).
36
Desirable properties of ScintillatorsIn
– High conversion efficiency
– Decay times of excited states should be short
– Material transparent to its own emissions
– Color of emitted light should match spectral sensitivity
of the light receptor
– For x-ray and gamma-ray detectors, µ should be large
– high detection efficiencies
– Rugged, unaffected by moisture, and inexpensive to
manufacture
37
Different materials for scintilators
• Sodium iodide activated with thallium
[NaI(Tl)], coupled to PMTs and operated in
pulse mode, is used for most nuclear
medicine applications
– Fragile and hygroscopic
• Bismuth germanate (BGO) is coupled to
PMTs and used in pulse mode as
detectors in most PET scanners
38
Photons Spectrometer
• An incident photon can deposit its full energy by:
– A photoelectric interaction (A)
– One or more Compton scatters followed by a
photoelectric interaction (B)
• A photon will deposit only a fraction of its energy
if it interacts by Compton scattering and the
scattered photon escapes the detector (C)
– Energy deposited depends on scattering angle, with
larger angle scatters depositing larger energies
39
Interactions of photon spectrometers
• Detectors normally shielded to reduce effects of
natural background radiation and nearby
radiation sources
• An x-ray or gamma-ray may interact in the shield
of the detector and deposit energy in the
detector:
– Compton scatter in the shield, with the scattered
photon striking the detector (E)
– A characteristic x-ray from the shield may interact with
the detector (F)
40
Spectrum of Cesium-137
• Cs-137 decays by beta particle emission to Ba137m, leaving the Ba-137m nucleus in an
excited state
• The Ba-137m nucleus attains its ground state by
the emission of a 662-keV gamma ray 90% of
the time
• In 10% of decays, a conversion electron is
emitted instead, followed by a ~32-keV K-shell
characteristic x-ray