Clinical Microscopy Procedures

13 flashcardsby MARIA
Study these flashcardsFree · spaced repetition · iPhone, Android, Mac & web

All cards

1
How does chemical tests or dipstick work?

1. Take the reagent strip and dip directly in the urine.
2. Compare the colors of the individual pads against the standards that are provided.
3. Wipe of excess fluid and wait a predetermined length of time.
11-3-2

1. Take the reagent strip and dip directly in the urine.
3. Wipe of excess fluid and wait a predetermined length of time.
2. Compare the colors of the individual pads against the standards that are provided.
2
How to calibrate SD Urometer 720 Urine Analyzers?

1. Normal calibration values are between 333+/-10 (323 to 343).
2. Result sprinted through thermal printer
3. Enter into "System Set Up" by pressing 'Esc' key twice
4. Select "System Configuration" and press 'ent' key3.
5. Press 'Esc' key to go back to Menu
6. Put the calibration strip on the strip load plate
7. Navigate to " White calibration" and "YE" press 'ent' key
3-4-7-5-6-2-1

3. Enter into "System Set Up" by pressing 'Esc' key twice
4. Select "System Configuration" and press 'ent' key
7. Navigate to " White calibration" and "YE" press 'ent' key
5. Press 'Esc' key to go back to Menu
6.Put the calibration strip on the strip load plate
2.Result sprinted through thermal printer
1.Normal calibration values are between 333+/-10 (323 to 343).
3
How to carry out the test in SD Urometer 720 Urine Analyzers?

1. 100 seconds later, result of first trip appears and next trip every 5 seconds
2. Put the test strip on to the strip loading plate after dipping in urine
2-1

2. Put the test strip on to the strip loading plate after dipping in urine
1. 100 seconds later, result of first trip appears and next trip every 5 seconds
4
How benedict's test is done?

1. Add 5 drops (0.25 ml) urine, then mix
2. With a tong place the test tube in boiling water bath for 5 minutes or boil over flame for 1 to 2 minutes
3. Place 5 ml of benedict's reagent into a test tube
4. Observe for the color change
3-1-2-4

3. Place 5 ml of benedict's reagent into a test tube
1.Add 5 drops (0.25 ml) urine, then mix
2. With a tong place the test tube in boiling water bath for 5 minutes or boil over flame for 1 to 2 minutes
4.Observe for the color change
5
How Exton's test/ Sulfosalicylic acid is done?

1. Grade the degree of turbidity using the chart
2. Observe cloudiness
3. Pipette 2 ml of exton's reagent and placed on the tube
4. Please for drops of yuri in the tube with exton's reagent
3-4-2-1

3. Pipette 2 ml of exton's reagent and placed on the tube
4. Please for drops of yuri in the tube with exton's reagent
2. Observe cloudiness
1.Grade the degree of turbidity using the chart
6
Steps in routine urinalysis (familiarize)
1. Lag the incoming urine specimen in the clinical microscopy and receipt and assign for accession number
2. Mix the urine gently by swirling and pour approximately 5 ml into the clean test tube.
3. Observe the color and transparency or appearance of the urine. Record it down.
4. Deep reagent strip into the urine sample, moistening all the pads .
5. Remove strip into the urine immediately and tap to remove the excess urine if possible blot edge on the absorbent tissue paper.
6. Put the test strip on to the strip loading plate. At the start time, number 5 appears on the LCD.
7. The number decreases and when it drops to 0, the system will start measuring. You can put the next strip when green light is on.
8. After 100 seconds, result of the first strip appears and next strip every after 5 seconds
9. Attach the urine chemistry result in your worksheet
10. Discard urine sample into the biohazard container
11.Centrifuge urine at 1500 to 2000 rpm for 5 minutes. Use a balance tube in every centrifugation.
12. Remove tube from centrifuge after the rotor spinning
13. Pour off supernatant urine, leaving approximately 0.5 mL of urine in tube
14. Resuspend urine sediment by tapping the bottom of the tube
15. Place one drop of resuspend urine into a clean glass slide
16. Play slide on the microscope stage and focus using low power (10x) and the condenser lowered
17. Scan 10 to 15 low-power fields, count the number of cast (if present) per field, record the average.
18. Rotate the high power objective (45x) into position (raise condenser if necessary). Scan 10 to 15 fields on high power
19. Count the number of wbc and rbc
20. Observe the sample for the presence of microorganism, crystals, mucus, epithelial cells, and estimate and record if present
21. If crystals are present identify type
22. Complete the urinalysis report
7
Stool Specimen Collection and Handling
• CoIIected in a clean con1ainer ( plastic or glass) with screw-capped lops.
• Must not be contaminated with urine or toiIet water.
• Examine liquid specimens within 30 minutes of passage.
• Examine soft specimens within 1 hour of passage.
• Examine formed stools within 24 hours.
8
Processing of Fecalysis (familiarize)
1. With a pencil or marker, write the patient's name or number and the left hand end of the slide.
2. Place a drop of saline in the center of the left half of the slide and place a drop of iodine solution in the center of the right half of the slides.
3. With the applicator stick, pick up a small portion of the specimen (size of a match head) and mix with the drop of saline.
4. In same process with the drop of iodine
5. Cover the drop of saline and drop of iodine with a coverslip.
6. Hold the coverslip at an angle, touch tye edge of the drop, and lower gently on to the slide.
7. Record the stool color and consistency
8. Put the slide with the mounts on the microscope stage and focus on the mount with the x10 or low-power objective.
9. Examin the entire coverslip area with the x10 objective
10. Record and print your results.
9
Fecal Occult Blood Test (FOBT) Processing (familiarize)
1. Using the applicator sticks provided, collect a small amount of stool specimen, on one end of the applicator. Apply a very thin smear in box 1.
2. Reuse applicator to obtain a second sample from a different part of the stool specimen. Apply a very thin smear inside box 2.
3. Allow the specimen to air dry, then close the cover.
4. Open perforated window on the back.
5. Apply two (2) drops of Hema-Screen Developer to the back side of the boxes 1 & 2.
6. Read results after 30 sec and within 2 mins.
7. Record the result any trace of blue color, within or on the outer rim of the specimen, is positive for occult blood.
10
Micral Test Processing
1. Dip the test strip into the urine for 5 seconds.
2. Place the strip on a nonabsorbed surface or across the top of collection cup.
3. Allow excess urine to drain and wait for 1 minute.
4. Compare the color of the detection pad on the strip with color scale on the test vial.
5. Record and print your result.
11
Pregnancy Test Processing

1. Record and print your result.
2. As the test begin to work, you will see a purple color band move across the result window in the center of the test device.
3. Interpret test result at 5 minutes
4. Holding the disposable dropper above the test device, squeeze 3-4 drops of serumrine into the sample well.
5. Remove the test device from the foil pouch, and place it on a flat, dry surface.
5-4-2-3-1

5. Remove the test device from the foil pouch, and place it on a flat, dry surface.
4. Holding the disposable dropper above the test device, squeeze 3-4 drops of serumrine into the sample well.
2. As the test begin to work, you will see a purple color band move across the result window in the center of the test device.
3. Interpret test result at 5 minutes
1. Record and print your result.
12
Sperm Macroscopic Examination (familiarize)
1. Record the date and time of collection, medication taken, and time of analysis.
2. Record the liquefaction time
- Take note if sample is already liquefied upon receipt.
- If the sample does not liquefy after 60 minutes, report time as > 1 hour.
3. Determine the viscosity
- Genty aspirate the sample into a wide-bore (approximately 1.5mm diameter) plastic disposable pipette allowing the semen to drop by gravity.
- And observing the lenght of any thread or introduce a glass rod into the sample and observing the length of the thread that forms upon the withdrawal of the rod.
4. Measure the volume
5. Determine the pH
- Mix the semen sample well
- Spread a drop of semen evenly on to the pH segment of the urine reagent strip
- Wait for the color of the impregnated zone to become uniform (<30 seconds)
- For normal samples, pH paper in the range 6.0 to 10.0 should be used
6. Record the pH.
13
Sperm Microscopic Examination (familiarize)
1. Making a wet preparation.
2. Examine the wet preparation using 100x magnification.
- Take note of sprem aggregation, degree of agglutination and point of attachment.
- Look for cells other than spermatozoa and estimate
3. Examine wet preparation using magnification x200 to x400
- Asses sperm motility
- Look for spermatozoa in an area at least 5mm from the edge of the coverslip
- Systematically scan the slide to avoid repeatedly viewing the same area
- Start scoring a given field at a random distant. Scan and count quickly.
- Tally the number of spermatozoa in each motility category
- Asses 200 spermatozoaeplicate
4. Aspirate semen up to 0.5 mark of WBC pipette.
- Aspirate diluting fluid up to 11.0 mark
- Shake horizontally to mix
- Charge the hemocytometer chamber, stand for 10 min to allow immobilized sperm settle.
- Count the sperm in 2 large squares (WBC squares)
- Calculate for the sperm concentration based on the following formula:
Sperm concentration (10x^6 per mL) number of sperms counted x 100,000
= 133 x 10^6 per mL x 100,000
= 13.3 x 10^6 per mL
- To get the total sperm count, multiply the sperm concentration by the volume of the whole ejaculate.

Total sperm count = sperm concentration x voulme
= 13.3 x 10^6 /mL x 2 ml
= 26.6 x 10^6 /ejaculate
5. Record and print your result.