1Which factors increase Compton scatter production?
Higher kVp; larger field size; greater patient thickness
2What happens to scatter production when x-ray field size increases?
Increases
3What happens to scattered radiation when the beam is collimated to a smaller field?
Decreases
4What happens to scatter production as kVp increases?
Increases
5How does increased scatter generally affect radiographic image contrast?
Decreases contrast
6Which type of contrast has few shades of gray and a short grayscale?
High or short-scale contrast
7Which window width corresponds to higher displayed image contrast?
Narrow window width
8Why can anatomical compression reduce scatter during imaging?
It reduces tissue thickness
9What does tissue compression do to object-to-image distance (OID)?
Decreases OID
10How does increasing the atomic number of the irradiated material affect scatter?
Decreases scatter as photoelectric absorption becomes more likely
11What two common radiopaque contrast substances are used in radiography?
Iodine; barium sulfate
12What does subtraction imaging remove to emphasize injected contrast material?
Background anatomy
13What is the most common beam-restricting device used in general radiography?
Collimator
14Name the three basic beam-restricting devices.
1. Aperture diaphragm; 2. Cone/cylinder; 3. Collimator
15A flat lead sheet with a central opening attached to the x-ray tube describes which beam restrictor?
Aperture diaphragm
16What is the principal drawback of an aperture diaphragm?
Increased penumbra (geometric unsharpness)
17What does penumbra mean in radiographic imaging?
Geometric unsharpness along an image edge
18How can penumbra associated with an aperture diaphragm be reduced?
Use a longer SID
19What is off-focus (extrafocal or stem) radiation?
Radiation produced away from the intended focal spot
20A circular aperture diaphragm with an extended metal tube describes which beam restrictor?
Cone or cylinder
21Which beam restrictor is more effective at reducing penumbra?
- Aperture diaphragm
- Flared cone
- Cylinder
- Lead mask
22What characteristic allows a collimator to adapt to many different examinations?
Adjustable field size
23What material forms the shutters of a radiographic collimator?
Lead
24Which collimator shutters primarily reduce off-focus radiation?
Upper shutters
25Which collimator shutters primarily reduce penumbra?
Bottom shutters
26What does the collimator light-localizing system use to project the approximate x-ray field?
Light source and mirror
27What does PBL stand for in beam restriction?
Positive Beam Limitation
28What is another term for a positive beam limitation device?
Automatic collimator
29What happens when a cassette or imaging plate enters a Bucky equipped with automatic collimation?
The light field is automatically adjusted to the receptor size
30Why should the light field not extend beyond the image receptor?
Unnecessary patient irradiation and image degradation
31What type of filtration does the collimator contribute?
- Inherent
- Compensating
- Total only
- Added
32Which two ancillary beam-restricting devices are discussed?
Lead blocker; lead mask
33What is a lead blocker typically made of?
Lead-impregnated rubber sheet
34Where is a lead mask typically attached for a specific field shape?
End of the collimator
35Why may a lead blocker help with a lateral lumbar-spine radiograph?
It absorbs scatter originating in posterior soft tissue
36Which beam-limiting device combines adjustable lead shutters and a projected light field?
- Collimator
- Aperture diaphragm
- Fixed cone
- Lead mask
37True or False: Opening the collimator field wider generally reduces scatter reaching the receptor.
False. A wider field generally increases scatter
38Which two physical factors determine the volume of irradiated tissue?
Field size; patient thickness
39Which tissue produces more scatter: bone or soft tissue?
Soft tissue
40In which imaging modality is anatomical compression especially common?
Mammography
41What is a major patient-dose benefit of restricting the x-ray beam to the anatomy of interest?
Less tissue is irradiated
42Which three advantages are associated with aperture diaphragms?
Simple design; low cost; ease of use
43For which applications are aperture diaphragms particularly suitable?
Chest radiography; fixed trauma units; dental equipment
44What accessory permits adjustment of an otherwise fixed cone field size?
Extension sleeve
45How does a lead blocker differ from a lead mask in typical placement?
A blocker is placed on the table, whereas a mask is secured to the collimator
46Which patient group benefits especially from lead blockers because of increased scatter production?
Larger patients