1Why are basic gradient-echo and spin-echo imaging slow, and what do fast sequences change?
In basic GRE and SE, each excitation (each TR) fills only one line of k-space, so you need as many TRs as there are phase-encode lines.
Fast sequences record several lines, or all of k-space, after a single excitation, by forming many echoes and giving each one a different phase-encode step.
2
How does fast spin echo (RARE, turbo spin echo) work?
One 90° pulse is followed by a train of 180° pulses, giving a train of spin echoes in the same TR.
Each echo gets its own phase-encode step, so each echo fills a different k-space line. After each echo the phase encoding is undone, so the next echo starts again from the centre row.
With an echo train of, say, 8 echoes, k-space fills 8 times faster than with one line per TR.
3
Fast spin echo: which echo sets the contrast, and what is the cost?
The echoes get weaker along the train because of T2 decay. The echo used for the centre of k-space sets the image contrast, so its time is called the effective TE.
Costs: lines recorded late in the train are weaker (T2 decay across k-space), and the many 180° pulses deposit a lot of RF energy (SAR).
4
What is a segmented acquisition?
k-space is split into several parts (segments). Each excitation fills one segment with an echo train, and a few excitations (shots) together fill the whole grid.
It sits between one line per TR (slow) and all of k-space in one shot (fastest). Shorter echo trains mean less decay during each train.
5
How does echo planar imaging (EPI) fill k-space?
After a single excitation, the readout gradient switches back and forth rapidly while small phase-encode "blips" step up one row each time.
Each lobe of the readout gradient records one line, alternately left to right and right to left, so the trajectory zig-zags through the whole of k-space in one shot (single-shot EPI). It can be built on a gradient echo or on a spin echo (90°–180°).
6
EPI: what are its strengths and weaknesses, and what does segmented EPI change?
Strength: extremely fast, a whole 2D image from one excitation, which is why it is used for diffusion MRI and fMRI.
Weakness: the whole readout happens while the signal decays (T2*), so it is very sensitive to B0 inhomogeneity and the decay weights k-space unevenly.
Segmented EPI fills k-space in several shots, each with a shorter echo train: less decay per train, but more excitations and so more time.
7
GRE vs SE vs fast spin echo vs EPI: how many k-space lines per excitation?
Gradient echo and spin echo: one line per TR.
Fast spin echo (RARE): one line per echo, several echoes per TR (echo train).
Segmented EPI or segmented FSE: one segment per shot, a few shots per image.
Single-shot EPI: the whole of k-space after one excitation.
8
Cell cultures vs organoids vs in vivo: what is each good for?
Cell cultures: basic cellular and metabolic questions.
In vivo: needed for complex organ function, because living tissue works as a whole, not as the sum of its components.
Organoids are an intermediate step. They can be kept alive in a bioreactor inside the same preclinical scanner and measured with MRS.
9
Preclinical RF coil set-ups: surface transmiteceive vs volume transmit with an array receive
Surface coil for both transmit and receive: very high SNR and B1 efficiency near the coil, but the B1 field (and so the flip angle) is very uneven with depth. Adiabatic pulses can make the flip angle insensitive to this.
Volume coil to transmit plus an array of small surface coils to receive: the volume coil gives an even flip angle everywhere, and the receive array gives high sensitivity.
10
What does ultra-high-field rodent 1H MRS make possible?
Higher SNR and spectral resolution than human MRS, so more than 18 metabolites can be measured in the rodent brain, some at concentrations down to about 0.5 mM (e.g. ascorbate, glycine, glutathione).
The better resolution helps to identify resonances and validate quantification, and to explore less-studied parts of the spectrum (e.g. downfield). Because it is non-invasive, the same animals can be followed over time (development, disease progression).
11
Why is 1H-MRSI not widely used in rodents?
1) Shimming is hard over a small brain with many tissue interfaces.
2) SNR is limited because the sample is small.
3) A lot of muscle, skin and fat lies within the coil's field of view, which causes water suppression artefacts and lipid contamination.
Fast 1H-FID-MRSI with saturation bands is the approach used to get around this.
12
Brain energy metabolism: which technique measures which step?
Glucose uptake from blood into tissue: FDG-PET, 2H MRS.
Glycolysis to pyruvate and lactate: 13C MRS, 1H functional MRS (lactate).
Acetate (glial) metabolism: 11C-PET, 13C MRS.
TCA cycle and glutamate–glutamine labelling: 13C MRS, 2H MRS.
Energy (ATP, PCr): 31P MRS.
Nitrogen metabolism (glutamine): 15N MRS.
13
How can 2H labelling be detected indirectly with 1H-FID-MRSI?
After infusing deuterium-labelled glucose, deuterium replaces some protons in its products, e.g. at the C4 position of glutamate and glutamine (Glx).
Deuterium does not give a 1H signal, so the 1H peak of those positions shrinks. Mapping the drop in the Glx C4 signal over time with 1H-FID-MRSI shows glucose oxidation, without needing a 2H coil.
14
How does dynamic 13C MRS give metabolic fluxes?
13C-labelled glucose (or acetate) is infused and 13C spectra are recorded repeatedly, so you see the label appear over time in different carbon positions (e.g. glutamate C4, glutamine C4).
These time courses are fitted with a metabolic model of glia and neurons, which gives the rates (fluxes) of e.g. the TCA cycle and the glutamate–glutamine cycle.
15
Rat practical: how was B0-map shimming done, and what does the field SD show?
A B0 map of the brain was acquired, a shim region (an ellipsoid) was drawn on it, and the scanner calculated the shim currents that best cancel the field distortion in that region.
The quality measure is the standard deviation (SD) of the field across the region, in Hz. A smaller SD means a more uniform field. In the practical it dropped from about 98 Hz to about 47 Hz, so the field became about twice as uniform.
16
Why is the shim region normally bigger than the voxel, and why not with a very large voxel?
Shimming a region slightly larger than the voxel makes the field uniform throughout the voxel, including at its edges.
With a very large voxel (as used for spectroscopic imaging), a bigger shim region would reach outside the brain into skull, air and fluid spaces. These have very different magnetic susceptibility, so the field there is very distorted and the shim cannot fit it, which makes the result inside the brain worse.
17
How does the field SD from the B0 map relate to the water linewidth?
Both describe how spread out the field is: a wider spread of field values gives a wider spread of frequencies, so a broader water peak. A smaller SD should give a narrower line, and the two are of similar size.
The SD is computed over the shim region. If that region is bigger than the voxel, the SD can be larger than the voxel's linewidth. After re-shimming on the voxel itself, the SD (about 28 Hz) and the linewidth (about 31 Hz) were close.
Running the shim again and getting almost the same SD means it has converged: that is the best this shim can do.
18
Is a water linewidth of about 31 Hz at 14.1 T good?
It is at the upper limit. What counts as good depends on field strength and voxel size.
Field inhomogeneity scales with B0, so the linewidth in Hz is larger at higher field: for a large voxel, roughly 16–20 Hz at 9.4 T but 24–29 Hz at 14.1 T. A smaller voxel should give a linewidth a few Hz narrower.
A broader line means lower peaks, so lower SNR (in the practical, a 26 Hz spectrum looked clearly better than a 32 Hz one), and poorer water suppression.
19
Why must the water frequency be set precisely after shimming, before water suppression?
Shimming changes B0, so it shifts the Larmor frequency. Water suppression uses narrow-band pulses that only act on a small frequency range, so they must sit exactly on the water peak.
If the frequency is slightly off, or the shim is poor so the water peak is broad with "shoulders", part of the water lies outside the suppression band and a large residual water peak remains.
20
Why were there dark spots where the surface coil touched the rat's head?
Right next to a surface coil the B1 field is very strong, so the flip angle there is far larger than intended, well beyond 90° (towards 180°).
A pulse near 180° leaves almost no Mxy, so those spots give very little signal and look dark, even though tissue is there.
21
Why does a tumour appear bright after injecting a contrast agent?
The contrast agent (e.g. gadolinium-based) shortens T1 of the water near it, so that tissue recovers Mz faster and looks bright on a T1-weighted image.
Normally it cannot cross the intact blood–brain barrier. In tumours such as glioblastoma the barrier is leaky, so the agent enters the tumour, which then stands out as a bright spot.
22
What is isotropic resolution, and what does it change?
Isotropic means the voxels have the same size in all three directions (e.g. 0.2 × 0.2 × 0.2 mm).
The image can then be viewed or resliced in any plane with the same detail, and tissue mixing within a voxel is the same in every direction.
Getting there usually means thinner slices, and since signal is proportional to voxel volume, thinner slices cost SNR.
23
Why do older rats give better image contrast than very young ones?
Contrast between grey and white matter comes largely from myelin. The brain of a very young rat is not yet fully myelinated, so grey and white matter look more alike, and contrast increases as myelination develops.
A small brain also means a fixed voxel size can be too big, mixing neighbouring regions (e.g. hippocampus and thalamus) or reaching outside the brain.