1When should a microcatheter be used for wiring?
2It should be considered in any complex lesion, such as highly tortuous lesions, near-occlusive lesions, and chronic total occlusions, and whenever more than one wire is expected to be needed. ELI5: A thin support tube around the wire gives better steering and lets you swap wires without starting over.::
What are the drawbacks of using an over-the-wire balloon in place of a microcatheter?
3It is stiffer and kinks more easily, and on small balloons the marker sits in the middle, so the position of the tip is hard to judge, although it costs much less. ELI5: It is a cheaper stand-in that bends badly and does not show you where its nose is.::
What are the wire options for a tortuous vessel?
4An advanced workhorse wire with a composite core, dual core, or nitinol core is tried first, then a soft-tip wire with hydrophilic coating, then a soft non-tapered polymer-jacketed wire. ELI5: Start with a wire that steers well, and move to progressively slipperier ones only if it fails.::
What is the hazard of polymer-jacketed, hydrophilic, or high tip load wires, and what is done after they cross?
5They carry a higher risk of dissection and distal perforation, so they are exchanged through a microcatheter for a workhorse wire before balloons and stents are delivered. ELI5: The slippery or stiff wire that gets through a hard lesion is also the one most likely to poke through the artery, so swap it out once it has done its job.::
How does the first wire choice for a chronic total occlusion differ from other lesions?
6The initial wire is usually a polymer-jacketed, tapered-tip wire, whereas other lesions start with a workhorse wire. ELI5: A blocked artery needs a thin, slick tip to find a microchannel, which you would not risk in an open artery.::
Which wire must be in place before atherectomy?
7The dedicated atherectomy wire for the device is required, a Rotawire for rotational atherectomy or a ViperWire for orbital atherectomy, placed by exchange through a microcatheter. ELI5: The drill only runs on its own special rail.::
How large should the bend on a guidewire tip be?
8The distal bend should be smaller than the diameter of the target vessel, so larger bends suit large vessels and small bends suit small vessels. ELI5: The hook has to fit inside the pipe to turn freely.::
When is a double bend useful on a guidewire tip?
9A larger proximal bend with a smaller distal bend improves reach in large vessels and vessels with angulated ostia. ELI5: The big curve gets you across a wide artery and the small curve does the fine steering.::
What is the rule for how much bend to put on a wire at first?
10A small bend is made first and increased only if needed, because bend can always be added but is hard to remove. ELI5: You can bend a wire more, but you cannot easily un-bend it.::
What is the specific precaution when shaping a nitinol wire?
11Shaping through an introducer can fracture the core, so a nitinol wire such as the Runthrough should be shaped only with the needle supplied with it. ELI5: This type of wire snaps inside if bent over the wrong tool.::
What can happen when a wire is advanced through a side-hole guide catheter?
12The wire can exit through the side hole instead of the catheter tip. ELI5: The wire takes the side door instead of the front door.::
How can a wire be advanced to the guide tip with little fluoroscopy, and what stops it from going too far?
13Shaft markers at 90 and 100 cm on some wires show when the tip is near the end of the guide, and a torquer tightened about 90 cm from the tip prevents over-advancement. ELI5: Marks on the wire tell you how far in it is, and a clamp acts as a stop.::
A wire has inadvertently entered the coronary artery. What should be done?
14It should not be pulled out, and contrast is injected to check where it is before redirecting it if necessary. ELI5: It may already be where you want it, so look before you undo it.::
How is a second wire passed safely when a guide extension, balloon, or stent is already in the guide?
15The other devices are removed if possible, and otherwise the wire is advanced under fluoroscopy without force or through the over-the-wire lumen of a dual lumen microcatheter. ELI5: Either clear the tunnel first or send the new wire down its own private lane.::
Which views are used to wire the left anterior descending and circumflex arteries?
16A caudal view is used to choose between them at the left main, then RAO cranial or AP cranial for the left anterior descending and RAO caudal or LAO caudal for the circumflex and marginals. ELI5: Look from below to pick the right fork, then from above for the front artery and from below for the back one.::
Which views are used to wire the right coronary artery?
17LAO is used to reach the distal vessel, and LAO or AP cranial to select the posterior descending or posterolateral branch. ELI5: One view for the main trunk and a tilted one to separate the two end branches.::
What should be done when the wire tip stops moving freely?
18The wire is withdrawn and readvanced, or left in place while a second wire is advanced alongside it, because pushing a tip that is under plaque causes dissection or entrapment. ELI5: A stuck tip is digging into the wall, so back out or go around it.::
When is a knuckled wire tip acceptable, and what are the limits?
19A small knuckle is safe and effective, especially after the lesion has been crossed, but it must stay within the radiopaque segment and should not be formed on highly supportive wires. ELI5: A small loop glides past side branches without poking into them, as long as it stays small and the wire is soft.::
How are multiple wires kept organized during bifurcation PCI?
20They are separated with a towel, with the wire in the upper branch on screen placed on top, or told apart by color, a bent back end, or a torquer. ELI5: Arrange the wires on the table the same way they look on the screen.::
The wire keeps entering the circumflex instead of the left anterior descending. What are the options?
21The tip can be reshaped, the guide repositioned or exchanged for a smaller curve, or a wire left in the circumflex to stabilize the guide while a second wire is steered into the target. ELI5: Change the aim of the wire or the catheter, or park one wire in the wrong branch as an anchor.::
How is the reversed or hairpin wire technique performed?
22A polymer-jacketed wire is bent about 3 cm from its tip, advanced folded into the vessel beyond the branch, and then pulled back so the tip drops into the sharply angulated branch. ELI5: Fold the wire into a hook, push it past the turn, and pull back so the hook catches the branch.::
What is the deflection balloon technique, and what are its drawbacks?
23A balloon is inflated in the main vessel just beyond the branch origin so that wires and microcatheters are deflected into the branch, at the cost of ischemia during the inflation and a risk of vessel injury. ELI5: Block the straight road so the wire has to take the turn, accepting that blood flow is blocked too.::
What is a pseudolesion, and how is it distinguished from dissection?
24A wire straightening a tortuous vessel creates folds that look like new stenoses, and partly withdrawing the wire so that only its soft part lies across the segment makes them disappear. ELI5: A stiff wire bunches the artery up like a sleeve, and easing the wire back smooths it out.::
The vessel closes after wire crossing and dissection is suspected. What must be confirmed before ballooning, and how?
25The wire must be confirmed in the distal true lumen, using intravascular ultrasound if needed, and contrast must not be injected distally through a microcatheter because it can extend a subintimal dissection. ELI5: Make sure you are in the real channel before inflating anything, and do not squirt dye into what might be the wall.::
What are the options when the wire is subintimal and the true lumen cannot be rewired?
26Reentry into the distal true lumen can be attempted with the Stingray system or penetrating wires, which requires chronic total occlusion expertise. ELI5: Use the tools designed for blocked arteries to steer back from the wall into the channel.::
How is a distal guidewire perforation managed?
27A balloon is inflated proximal to the perforation to stop bleeding into the pericardium, followed by fat or coil embolization if extravasation continues. ELI5: Block the artery upstream of the leak first, then plug the tiny branch permanently.::
When is distal wire perforation most likely to go unnoticed?
28It is most likely when collimation hides the wire tip, when several wires are in use, and when difficult delivery or withdrawal of equipment moves the wire. ELI5: The tip creeps forward while you are looking elsewhere or wrestling with equipment.::
Why is losing wire position proximal to a freshly treated lesion dangerous?
29If dissection closes the vessel there may be no way back across it, so equipment should not be forced and the lesion should be better prepared instead. ELI5: The wire is your lifeline through the lesion, and shoving a stuck stent can pull it out.::
What are the four ways to remove an over-the-wire device over a short guidewire?
30Trapping, hydraulic exchange, a guidewire extension, and cutting the device off in segments are the four methods, and trapping is preferred. ELI5: Pin the wire, push it with fluid, lengthen it, or cut the tube off it.::
Why is trapping the preferred exchange technique?
31It minimizes wire movement, which protects against loss of position and distal perforation, and it needs little fluoroscopy. ELI5: The wire is pinned still inside the guide, so nothing shifts at the far end.::
What size, length, and pressure are used for a trapping balloon?
32A 2.5 mm balloon is used in 6 or 7 French guides and a 3.0 mm balloon in 7 or 8 French guides, ideally 20 mm or longer, inflated to 15 to 20 atm. ELI5: The balloon has to fill the guide tightly and grip a long stretch of wire.::
Where is the trapping balloon positioned?
33It is advanced alongside the wire, not over it, to just beyond the tip of the withdrawn microcatheter inside the guide, usually near the primary curve. ELI5: The balloon goes next to the wire, ahead of the tube you are removing, so it pins bare wire against the guide wall.::
What step must never be skipped after deflating a trapping balloon?
34The Y-connector must be back bled, because air is often drawn into the guide during trapping and would be injected into the coronary artery. ELI5: Let the air out before you inject, or you will send a bubble into the heart.::
Which wires are hardest to trap, and why?
35Polymer-jacketed wires are slippery and can slide past the inflated balloon, so a longer balloon and gentle withdrawal are needed. ELI5: A slick wire can slip out of the grip.::
What are the limits of trapping in a 6 French guide?
36Over-the-wire balloons, the CrossBoss, and the Stingray balloon cannot be trapped, although low-profile microcatheters can be trapped in 6 French guides with a 0.071 inch inner diameter. ELI5: A small guide has no room for a trapping balloon next to bulky devices.::
How is hydraulic exchange performed?
37A saline-filled inflation device is connected wet to wet to the microcatheter hub and held at 14 to 20 atm while the microcatheter is withdrawn under fluoroscopy. ELI5: Fluid pressure pushes the wire forward as fast as you pull the tube back, so the wire stays put.::
When should hydraulic exchange be avoided?
38It should be avoided with stiff, tapered, or polymer-jacketed wires and in chronic total occlusion PCI because the wire can be driven forward and perforate, and with over-the-wire balloons and the Stingray because friction pulls the wire back. ELI5: The fluid push is imprecise, which is unacceptable when the wire is sharp or the device grips it.::
What is the main practical caveat with guidewire extensions?
39Each wire brand needs its own matching extension, and the connection must be tightened firmly without kinking the wire. ELI5: The add-on piece only fits its own make of wire and must be screwed on tight.::
When is the circumcision technique useful?
40It is useful when the wire is stuck inside the microcatheter, or when trapping is not possible and no matching extension is available, and it destroys the device. ELI5: When nothing else works, cut the tube off the wire piece by piece.