Hemostasis and Coagulation Instrumentation (Rodak's Hematolo

118 flashcardsby karebear
Study these flashcardsFree · spaced repetition · iPhone, Android, Mac & web

All cards

1
Q: What tests made up the traditional coagulation menu?
A: PT/INR, APTT, fibrinogen, thrombin time
2
Q: How did coagulation testing evolve with new instrumentation?
A: Specialized tests became routine, testing became standardized, consistent, and cost-effective
3
Q: What is still required despite automation?
A: Method validation, cognitive ability, and theoretical knowledge of hemostasis
4
Q: What are the six main end-point detection principles?
A: Mechanical, photo-optical, nephelometric, chromogenic, immunologic, viscoelastic
5
Q: How does electromechanical clot detection work?
A: Measures change in conductivity between two electrodes; clot completes the circuit and stops the timer
6
Q: What device used electrodes for clot detection?
A: BBL Fibrometer
7
Q: How do steel ball methods work?
A: A ball oscillates in plasma; fibrin increases viscosity and slows movement until timer stops
8
Q: Which analyzers use oscillating ball methodology?
A: Diagnostica Stago analyzers
9
Q: How does photo-optical detection work?
A: Measures changes in plasma optical density (light transmittance) during clotting
10
Q: What interferes with photo-optical methods?
A: Lipemia, icterus, hemolysis, or translucent clots
11
Q: How do modern instruments minimize interference?
A: Use multiple wavelengths to filter effects of icterus/lipemia
12
Q: What substrate is typically used in chromogenic assays?
A: Synthetic peptide substrate with para-nitroaniline (pNA)
13
Q: How does the chromogenic method measure enzyme activity?
A: Cleavage releases yellow pNA, OD at 405 nm is proportional to protease activity
14
Q: How does the chromogenic method measure enzyme activity?
A: Cleavage releases yellow pNA, OD at 405 nm is proportional to protease activity
15
Q: Difference between direct and indirect chromogenic assays?
A: Direct = OD proportional to analyte activity; Indirect = OD inversely proportional due to inhibition of a target enzyme
16
Q: How does nephelometry work?
A: Measures light scatter at 90° or forward-angle as fibrin polymers form
17
Q: What kind of result does nephelometry provide?
A: A clot curve or “signature” when continuous readings are taken
18
Q: What is immunologic end-point detection based on?
A: Antigen–antibody reactions with latex microparticles
19
Q: How is light absorbance used in immunologic assays?
A: Latex-coated microparticles agglutinate when bound to antigen, increasing light absorbance
20
Q: Example of analyte measured by immunologic methods?
A: D-dimer
21
Q: What does viscoelastic clot detection measure?
A: Time to clot, clot strength, clot kinetics, and fibrinolysis
22
Q: Which instruments use viscoelastic methods?
A: TEG (Thromboelastography) and ROTEM (Rotational Thromboelastometry)
23
Q: Advantage of mechanical clot detection?
A: Not affected by lipemia or icterus; detects weak clots
24
Q: Disadvantage of photo-optical detection?
A: Interference from lipemia, icterus, and hemolysis
25
Q: Advantage of chromogenic assays?
A: Measure proteins that do not clot; specific
26
Q: Disadvantage of immunologic methods?
A: High cost and limited automated tests available
27
Q: What are the three levels of coagulation instrumentation?
A: Manual, semiautomated, automated
28
Q: Describe manual coagulation instruments.
A: Reagents/specimens transferred manually; temperature via water bath/heat block; end-point detected visually by operator
29
Q: Examples of manual methods?
A: Tilt tube, wire loop
30
Q: Describe semiautomated coagulation instruments.
A: Manual transfer of reagents/specimens; internal 37°C temperature control; automatic timing and clot detection
31
Q: Examples of semiautomated analyzers?
A: Fibrometer, STart 4, Cascade M & M-4, BFT-II, KC1, KC4
32
Q: Examples of semiautomated analyzers?
A: Fibrometer, STart 4, Cascade M & M-4, BFT-II, KC1, KC4
33
Q: Describe automated coagulation instruments.
A: Reagents pipetted automatically; internal monitoring of 37°C; timers and clot detection automatic
34
Q: Examples of automated analyzers?
A: ACL TOP, STA-R Evolution, STA Compact, Sysmex CA-series, BCS XP
35
Q: Why were duplicate tests required in visual methods?
A: Coefficient of variation >20%
36
Q: What precision is achieved by automated analyzers?
A: CV <5%, sometimes <1%
37
Q: What is random access testing?
A: Ability to run multiple tests in any order on one/many specimens
38
Q: Advantages of random access testing?
A: Faster turnaround, reduced errors, reduced labor
39
Q: How have reagent and specimen volumes improved?
A: Tests now require half or one-quarter of traditional volumes
40
Q: Why are open reagent systems important?
A: Allow labs to choose reagents from different manufacturers
41
Q: What features support reagent tracking?
A: Lot/expiration records, volume monitoring, barcoding
42
Q: What is primary tube sampling?
A: Analyzer aspirates directly from centrifuged collection tube
43
Q: Benefits of closed-tube sampling?
A: Safer, reduces exposure, stabilizes pH, prevents crossover
44
Q: What quality step is needed with closed-tube sampling?
A: Visual check for fibrin strands/clots after centrifugation
45
Q: How do analyzers handle dilutions?
A: Perform automatic dilutions of specimens, calibrators, and controls
46
Q: What features do modern analyzer computers include?
A: Data storage/ retrieval, calibration curves, QC files, Westgard rules, automatic repeat testing
47
Q: What is bidirectional LIS interface?
A: Analyzer queries LIS for ordered tests and sends results back
48
Q: What is reflex testing?
A: Automatic additional testing/dilutions if results exceed preset limits
49
Q: What is a clot “signature”?
A: Graph of clot formation kinetics showing baseline, acceleration, deceleration, and end-point
50
Q: What does a “clotted” flag mean?
A: Shortened clotting times due to premature activation
51
Q: How does lipemia affect results?
A: Falsely prolongs clotting times on optical instruments
52
Q: How does hemolysis affect results?
A: Falsely shortens clotting times via factor/platelet activation
53
Q: How does icterus affect results?
A: Prolongs clotting times (factor deficiency) and interferes with OD instruments
54
Q: What does “no end-point detected” mean?
A: Instrument couldn’t detect clot; alternative method needed
55
Q: What factors should guide the selection of coagulation instruments?
A: Patient population needs, test menu, cost-effectiveness, and standardization
56
Q: What might be appropriate for a small hospital with low-volume esoteric testing?
A: A batch analyzer with high throughput for routine coagulation tests
57
Q: Name some criteria for selecting instruments (Box 42.3).
A: Cost, service response time, reliability, maintenance, ease of use, test menu breadth, LIS interface, reagent flexibility, training support
58
Q: What are examples of specialized features on current coagulometers (Table 42.3)?
A: Random access, primary tube sampling, cap piercing, barcoding, bidirectional LIS, flagging, liquid level sensing, onboard QC, stat capability, refrigeration, specimen storage, reflex testing, clot graphs
59
Q: What are POC coagulation instruments?
A: Handheld/ bedside devices for near-patient testing, surgeries, or self-testing
60
Q: Advantages of POC instruments?
A: Fast turnaround, portability, small specimen volume (useful in infants)
61
Q: What specimen is typically used in POC coagulation devices?
A: Capillary (fingerstick) whole blood
62
Q: How is testing usually performed on POC devices?
A: Blood applied to a cartridge with reagents, then inserted into the analyzer
63
Q: What was the first POC coagulation test introduced in 1966?
A: Activated clotting time (ACT) for heparin monitoring in cardiac surgery
64
Q: How is heparin therapy reversed after cardiac surgery?
A: With protamine, monitored by ACT
65
Q: What is the most common POC test today?
A: PT/INR for Coumadin (warfarin) monitoring
66
Q: Why must POC results be validated against central lab assays?
A: Because correlation can be weak due to differences in reagents, specimens, and instrumentation
67
Q: What INR results from POC testing require confirmation by central lab?
A: INR >6.0 or unexpected INR changes
68
Q: What are some POC instruments and their detection principles?
Abbott iSTAT: electrogenic
Alere INRatio: electrochemical
Helena Cascade POC: photo-mechanical
Accriva ProTime Micro: mechanical clot
Roche CoaguCheck XS: amperometric
Siemens Xprecia Stride: electrochemical (amperometric)
69
Q: What does global hemostasis assessment measure?
A: The entire kinetic process of whole blood clot formation (initiation, strength, fibrinolysis, hypercoagulability)
70
Q: What is the advantage of using whole blood in global assays?
A: Allows evaluation of platelets, RBCs, WBCs, coagulation factors, and proteins together
71
Q: When is thromboelastography (TEG) useful?
A: In hepatic disease, obstetrics, trauma, cardiac surgery, or liver transplant
72
Q: How does TEG work?
A: Whole blood in an oscillating cup (4.75°); fibrin links cup and stationary pin; pin torque transmitted to torsion wire → electrical signal → tracing
73
Q: How does ROTEM differ from TEG?
A: In ROTEM, the pin rotates while the cup is stationary; less vibration-sensitive, 4 channels, touch screen, automated pipettor
74
Q: What do TEG/ROTEM tracings show?
A: Clot initiation (R/CT), clot kinetics (K/CFT, α-angle), clot strength (MA/MCF), fibrinolysis (Ly30/CLI)
75
Q: Disadvantages of TEG/ROTEM?
A: Operator-dependent, require training, interpretation requires expertise
76
Q: Why is platelet function testing important?
A: To monitor antiplatelet therapy (aspirin, clopidogrel, GP IIb/IIIa inhibitors) and assess bleeding risk
77
Q: Why is bleeding time no longer used?
A: It’s technically demanding, operator-dependent, and not predictive of surgical bleeding
78
Q: What principle is used in classic platelet aggregometry?
A: Light transmission aggregation (LTA) – increased light transmission as platelets aggregate
79
Q: What parameters are measured in LTA?
A: Maximum % aggregation, area under the curve, velocity
80
Q: What is impedance aggregometry?
A: Measures platelet adhesion to electrodes in whole blood
81
Q: What is lumi-aggregation?
A: Uses luminescence to measure secretion + aggregation simultaneously
82
Q: What is the BioData PAP-8E?
A: An 8-channel platelet aggregometer using LTA, low sample volume (225 μL), with touch screen and barcode scanner
83
Q: What is the Chrono-Log 700 Lumi-Aggregation System?
A: Aggregometer for whole blood or PRP; uses impedance + luminescence detection
84
Q: What is the Multiplate Analyzer?
A: Whole blood impedance-based aggregometer; used for monitoring clopidogrel and GP IIb/IIIa inhibitor therapy
85
Q: What is the Helena AggRAM?
A: Modular platelet aggregometer with laser diode optics at 650 nm, 4 channels, microvolume testing
86
Q: What is the Siemens PFA-100?
A: Automated platelet function analyzer using whole blood through collagen/ADP or collagen/epinephrine membranes; closure time indicates function
87
Q: What conditions can the PFA-100 detect?
A: von Willebrand disease, aspirin effect
88
Q: What is the VerifyNow system?
A: Cartridge-based optical detection system using fibrinogen-coated beads and agonists; provides aspirin, GP IIb/IIIa, and P2Y12 assays
89
Q: What is Plateletworks?
A: Assay kit run on hematology cell counters; compares platelet counts before and after agonist activation
90
Q: What are advantages of flow cytometry for platelet function?
A: Whole blood testing, small volume (~5 μL), sensitive in thrombocytopenia, measures receptor upregulation, microparticles, platelet-leukocyte binding
91
Q: What platelet disorder can be diagnosed using flow cytometry?
A: Heparin-induced thrombocytopenia (HIT)
92
Q: What is molecular coagulation testing used for?
A: To detect thrombophilia-related mutations and inherited bleeding disorders
93
Q: What are the most common mutations tested in thrombophilia?
A: Factor V Leiden and Prothrombin G20210A
94
Q: Which gene mutation is also tested but has unclear clinical utility?
A: MTHFR (methylene-tetrahydrofolate reductase)
95
Q: What does ACMG recommend about MTHFR testing?
A: It should not be routinely performed in thrombophilia workups
96
Q: What is the most common molecular testing method?
A: PCR-based assays
97
Q: What mutations can PCR detect?
A: Point mutations and single-nucleotide polymorphisms
98
Q: What is RFLP analysis used for in coagulation testing?
A: To analyze PCR products, though not suitable for high-volume labs
99
Q: Name other PCR-based/non-PCR methods available.
A: ARMS (amplification refractory mutation system), LightCycler, array technology, Invader assays, ligand-based technologies
100
Q: What is an advantage of molecular testing compared to clot-based assays?
A: High sensitivity/specificity and no interference from anticoagulants/inhibitors
101
Q: What sequencing method has become the primary tool for detecting variants?
A: High-throughput sequencing (HTS)
102
Q: What are the advantages of HTS?
A: Parallel sequencing of millions of molecules, whole exome analysis, identifies mutations/variants
103
Q: What role does molecular testing play in hemophilia and VWD?
A: Identifies mutations in FVIII, FIX, and VWF genes; helps with diagnosis and classification
104
Q: What is a lateral flow assay?
A: A paper-based platform using antibody detection for rapid, low-cost results
105
Q: How does a lateral flow test strip work?
A: Specimen moves by capillary action through zones with reagents; antibody-conjugated particles bind target analyte and form visible lines
106
Q: What is a lab-on-a-chip (LOC)?
A: A microelectronic chip integrating fluid handling and protein assays using tiny blood volumes
107
Q: What is the advantage of lab-on-a-chip?
A: Simultaneous measurement of several proteins with increased sensitivity and rapid results
108
Q: How will future technologies help coagulation labs?
A: Provide low-cost, sensitive, user-friendly tools for diagnosing thrombotic/ bleeding disorders
109
Q: Which are examples of technology advances in coagulation labs?
A: Primary tube sampling, random access testing, reduced reagent/specimen volumes (all of the above)
110
Q: What is an advantage of mechanical end-point detection?
A: Not affected by lipemia
111
Q: What method of detection is used in both TEG and ROTEM?
A: Viscoelastic
112
Q: Which is a feature of semiautomated analyzers?
A: Manual addition of reagents/specimens; automatic timing/clot detection
113
Q: Which method is most susceptible to icterus interference?
A: Photo-optical clot detection
114
Q: Which platelet test method works for thrombocytopenia/pediatrics?
A: Flow cytometry
115
Q: Which is NOT a factor for analyzer selection?
A: Manufacturer’s home office location
116
Q: What method does the PFA-100 use?
A: Closure time as blood flows through aperture coated with platelet agonists
117
Q: Where is POC testing used in high volume?
A: Monitoring oral anticoagulant therapy
118
Q: Molecular testing can diagnose which disorders?
A: Factor V Leiden and Prothrombin G20210A mutations
Study all 118 cards for free

More flashcards

Coagulation and Thrombosis Part 3/3121 cardsMedicine & health · EnglishJESSQTRANUE 2.1.S1 Chapitre 5 : Les tissus85 cardsFrenchClaraSNU FALL22-2165 cardsEnglishClaytonCapitals of US States50 cardsHistory & humanities · Englishbtoff49Week 12 & 13 Spreadsheet Study Guide41 cardsComputer science · EnglishSgslovíčka 0-tá lekce33 cardsCzechZuzanaSize and Surface Area27 cardsScience · EnglishLoganVirtual Memory22 cardsComputer science · English—Griffith—stage 15 vocab terms18 cardsLanguages · Latin → EnglishAnielaC2007Geometrija 3D14 cardsMaths · DutchBorftPsychology: Sensation and Perception (Ch 5)10 cardsEnglishuzachihaspanish vocab168 cardsLanguages · Spanish → Englishmkv

Study these decks in the app

Free on iPhone, Android, Mac and the web. Add any deck to your account in one tap.

Download on the App StoreGet it on Google Play
Study all 118 cards for free