MRI Physics: Scanner Setup and Tuning

33 flashcardspar Moll
Réviser ces flashcardsGratuit · répétition espacée · iPhone, Android, Mac et web

Toutes les cartes

1
What is the order of the scanner set-up, step by step?
1) Safety checks; sample in the coil, coil in the magnet.
2) Tune & match the loaded coil.
3) Find the water frequency (FID → spectrum).
4) Calibrate RF power (which power gives 90°).
5) Localizer images.
6) Shim the region of interest, then re-centre the frequency.
7) Set receiver gain.
8) Run the protocol and check a first image.
2

Why does the set-up follow that order?
Each step needs the previous one to be right.
Coil coupling must work before any MR signal is reliable; the pulse must be on resonance before you can calibrate its flip angle; the localizer tells you where to shim; shimming can shift the frequency, so you re-centre after. (Practical 1 did localizer/shim before the RF calibration; automatic routines also reorder, but the dependencies stay the same.)
3

Practical 1: why was copper sulfate added to the water phantom?
Paramagnetic Cu2+ ions shorten water's T1 (and T2), so Mz recovers quickly and scans can be repeated fast without saturation.
Pure water has a T1 of seconds, which would force very long waits between repetitions.
4

Practical 1: why a volume coil for the phantom?
Its B1 is uniform over the whole cylinder, so every region gets the same flip angle and similar sensitivity.
A surface coil would give strong signal only near the loop and fading signal deeper in.
5

Practical 2: why was the brain in an "oil with no hydrogen"?
Most likely a proton-free fluorinated fluid (e.g., Fomblin): it gives no 1H signal (black background), keeps the tissue from drying, and reduces the susceptibility jump an air boundary would cause.
Ordinary oil contains hydrogen and would show up.
6

Tune & match: when do you do it and why?
First, with the sample already in the coil in its final position, before any MR measurement.
A tuned and matched coil turns RF power into B1 efficiently (correct flip angles) and passes the weak MR voltage to the receiver (good SNR).
7

How do you read the tune/match (wobble) curve?
x-axis: frequency around the target (≈ 600 MHz for 1H at 14.1 T); y-axis: reflected power.
Goal: a deep dip exactly at the target frequency. Practical criterion: < ~10% reflected (0% is ideal but hard to hold).
8

The dip is deep but sits off the target frequency. What do you adjust?
Tuning. It changes a capacitor, which shifts the coil's resonance f_coil = 1/(2π√(LC)) and slides the dip sideways onto the target.
9

The dip is at the target but shallow. What do you adjust?
Matching. It brings the coil's impedance to 50 Ω so less power bounces back, which deepens the dip.
Then re-check tuning: the two knobs interact, so you iterate.
10

Why did the wobble curve change when the sample or coil moved, without touching the knobs?
Sample loading: the sample's size, conductivity and position change the coil's effective L, C and losses, so its resonance and impedance shift.
Re-tune after repositioning or changing samples. A disconnected cable gives total reflection: the dip disappears.
11

Finding the water frequency: what do you actually do?
With gradients off: pulse, record the FID, Fourier-transform it, find the water peak, and set f_RF and f_ref onto it.
(γ/2π)·B0 is only the starting guess; drift and the local field move the true resonance slightly.
12

The water peak sits 1.5 kHz above zero on the spectrum. What do you do and what changes?
Raise f_ref by 1.5 kHz, so Δf = f_L − f_ref → 0.
The peak moves to 0 and the FID stops oscillating (smooth decay). The width stays the same: centring is bookkeeping, not a field correction. (Peak 30 Hz below zero → lower f_ref by 30 Hz.)
13

RF power calibration: what do you do and what do you look for?
Keep power fixed and step the pulse duration (or fix the duration and step the power); record the initial FID amplitude each time.
First maximum = 90°; back to ≈ 0 at 180°; maximum again at 270°. Signal ∝ |sin θ|. Why: the console must know what setting gives the flip angle you ask for in this coil + sample.
14

What must be kept constant between calibration trials?
Recovery time (full Mz each time: long TR or a short-T1 sample), receiver gain, and the delay before the amplitude is read.
Otherwise a change in amplitude could come from saturation or scaling instead of the flip angle.
15

Worked: 90° needs 10 W for 60 µs. What gives 180°?
120 µs at 10 W, or 40 W at 60 µs.
θ ∝ duration, so double the time; or θ ∝ B1 ∝ √P, so double B1 = 4× power. (Same coil, sample and pulse shape.) The practical sweep started at ~10 µs and found the first maximum near 60 µs.
16

Why was there still some FID after a nominal 180° pulse?
B1 is not perfectly uniform (and some spins are off resonance), so parts of the sample get more or less than 180° and keep some Mxy.
For the same reason the calibration minimum is shallow and the maxima are less sharp: the total signal mixes regions with different flip angles.
17

What is "reference power" in practical 2?
The power that gives a 90° flip with a standard reference pulse (here noted as 1 ms) in this coil + sample. Other pulses are scaled from it by angle, duration and shape.
Lower reference power = more efficient transmit for that set-up. For a transmiteceive coil that usually also means better receive sensitivity (reciprocity), but it is a transmit calibration, not an SNR measurement. If shown as attenuation in dB, more dB = less power.
18

Why acquire a localizer?
Quick, low-resolution GRE images in three orthogonal planes to check the sample is centred in the magnet and coil, and to plan slices, voxels and the shim region.
It is for planning; its short TR/TE/small flip are chosen for speed, not for a particular contrast.
19

What caused the dark cross in the localizer?
Where the three planes intersect, the same spins are excited by more than one plane's pulses, so they have less time to recover → less Mz → darker bands.
It is saturation, not missing sample or bad shimming. Stronger with long T1 or short spacing between pulses; fades when T1 is short.
20

Why can a uniform phantom look unevenly bright?
1) B0 inhomogeneity → extra dephasing (GRE especially; improves after shimming).
2) B1 non-uniformity in transmit or receive.
3) Air bubbles: no water plus a susceptibility jump, so the dark area is bigger than the bubble.
4) Saturation at slice intersections.
21

Shimming: when do you do it and what do you watch?
After the localizer has defined the region, with gradients off. Adjust shim currents until the water peak is narrow and smooth and the FID lasts longer.
Physics: a more uniform B0 → more similar Larmor frequencies → slower dephasing (longer T2*). Re-centre the frequency afterwards.
22

In what order are shims adjusted, and how do you judge them?
Linear (x, y, z) first, then second order (z^2, xy, …); go back and forth because the patterns overlap.
Judge by linewidth (FWHM) and FID persistence, not by peak height, which also depends on gain and flip angle.
23

Narrow peak off-centre vs broad peak at zero: what fixes each?
Narrow but off-centre → move f_ref (frequency centring). Broad but centred → shim (the field is non-uniform).
One fixes the electronic reference, the other fixes the actual field.
24

B0 map vs B1 map: which problem does each diagnose?
B0 map: where the field/frequency is off → guides shimming. B1 map: where the achieved flip angle is off → diagnoses RF coil non-uniformity.
Uneven brightness can come from either, so they are not interchangeable checks.
25

Receiver gain: what does it do and not do?
It amplifies the received voltage so the strongest signal fills the ADC range without clipping.
Too high → clipping; too low → digitization noise matters. It does not change B1, the flip angle or the spins, and cannot raise the sample's intrinsic SNR.
26

Lookalike knobs: what does each one physically change?
1) Tuning → coil resonance (dip position).
2) Matching → coil impedance / power transfer (dip depth).
3) Frequency centring → electronic reference (peak position; spins unchanged).
4) Shimming → the actual B0 field (peak width, FID length).
5) RF power → B1 amplitude (flip angle).
6) Receiver gain → amplification (signal scale, clipping).
Trap: tuning moves the coil's resonance, not the protons' frequency; centring moves the peak, not its width.
27

1D profile of the cylinder: across the diameter vs along the length
Across the diameter: a rounded profile, because central slabs cut through a wider chord of the circle → more water. Along the length: nearly rectangular, because every slab has the same circular cross-section.
The concentration is uniform; the shape shows how much water sits at each position.
28

Worked: the profile spans 8516 Hz with G = 0.020 T/m. How big is the sample?
D = Δf / ((γ/2π)·G) = 8516 / (42.58 × 10^6 × 0.020) = 0.010 m = 10 mm.
Across the tube this is the diameter; along it, the length.
29

Readout gradient made stronger, or reversed: what happens to the profile?
Stronger G → the profile spans more Hz, but converting back with the larger G gives the same size: the object did not change.
Reversed polarity → mirror image (a symmetric cylinder looks the same).
30

GRE vs SE echo amplitude as TE increases: what did you see and why?
Both fall with TE, but GRE falls faster: it decays with T2* (field spread not refocused), SE with T2 (static spread refocused by the 180°).
Fit exponentials to the echo-centre amplitudes; T2* ≤ T2, and better shimming brings GRE closer to SE.
31

What did opening the ADC early in the spin-echo demo show?
The signal rising as the spins rephase after the 180°, peaking at TE, then falling again as the static field differences dephase them once more.
So one echo's shape shows T2*-type dephasing on both sides, while echo-centre heights across different TEs give T2. Recording could only start after the dead time (coil ring-down, T/R switching).
32

Practical 2: what did TE = 9 ms and a long TR imply for contrast?
Little T2 contrast (short TE) and little T1 contrast (long TR) → mostly proton-density weighting.
Ex vivo or fixed tissue has different T1/T2 from living tissue, so the familiar in vivo brightness pattern may not appear: "short" and "long" are relative to the sample.
33

Which adjustments are reused between scans, and when must they be redone?
Frequency, shim and reference power are reused while the set-up is unchanged; redo them after a new sample or coil, repositioning, a new target region or drift. Receiver gain and method-specific adjustments run for each scan.
Before committing to a long scan, check a short one for wrap, clipping, noise and dark bands, plus time and SARradient limits.
Réviser les 33 cartes gratuitement

Plus de flashcards : Sciences

GCSE Edexcel physics 6.1-Atomic Structure26 cartesSciences · AnglaisKonradRadiation and Atoms22 cartesSciences · AnglaisLemEnergy resources18 cartesSciences · Anglais17Module 4: Elektrische systemen (deel 1)14 cartesSciences · NéerlandaisYaniFiltration, Electron–Anode Interactions & Emission Spectrum 10 cartesSciences · AnglaisHaleemaRadiology Full Exam 176 cartesSciences · AnglaisAndyBiochemistry: Carbohydrates103 cartesSciences · AnglaisPhilipGeneral Chemistry: Electronic Structure77 cartesSciences · AnglaissiaMixed Signals: Transducers and Sensors59 cartesSciences · AnglaisDiandrei4- cellules gliales du SNC48 cartesSciences · FrançaisOcéaneAstronomy 2022 November 21-2339 cartesSciences · AnglaisEmmachp 12: biotechnology and genomics32 cartesSciences · Anglaislizzieowizzie

Tous les decks : Sciences

Révise ces decks dans l’app

Gratuit sur iPhone, Android, Mac et le web. Ajoute n’importe quel deck à ton compte en un tap.

Télécharger dans l’App StoreDisponible sur Google Play
Réviser les 33 cartes gratuitement