MRI Physics: k-space, FOV, and Resolution

25 Karteikartenvon Moll
Diese Karteikarten lernenKostenlos · Spaced Repetition · iPhone, Android, Mac & Web

Alle Karten

1
What sets the field of view (FOV)?
The spacing between neighbouring k-space samples: FOV = 1/Δk. Smaller steps in k-space give a larger FOV.
Why: one step Δk adds exactly one full 360° cycle of phase offset across the FOV. Positions exactly one FOV apart therefore always have the same phase offset, so the scanner cannot tell them apart.
2

What sets the pixel size (resolution)?
How far out in k-space you sample. The total k-range W_k = N·Δk sets the pixel size: pixel = 1/W_k = FOV/N.
Sampling further out (higher spatial frequencies) gives finer pixels.
3

Worked: FOV 32 mm with 128 pixels
Pixel size = 32 mm / 128 = 0.25 mm.
Checking from k-space: Δk = 1/0.032 m = 31.25 cycles/m. Total range W_k = 128 × 31.25 = 4000 cycles/m, and 1/4000 m = 0.25 mm. Same answer from both sides.
4

How do the readout gradient and dwell time set FOV_x and resolution?
The dwell time Δt is the time between two readout samples. During it, the readout gradient adds a little net area, which is one k-space step: Δkx = γ̄·Gx·Δt, so FOV_x = 1/(γ̄·Gx·Δt).
The whole readout (N samples, time T = N·Δt) sets the total kx range, and therefore the pixel size. (γ̄ = γ/2π)
5

How do the phase-encode steps set FOV_y and resolution?
The difference in area between two neighbouring phase-encode steps sets the spacing: Δky = γ̄·ΔGy·τ, so FOV_y = 1/Δky.
The largest phase-encode step (how many steps you take) sets how far out ky goes, and therefore the pixel size in y.
6

More outer samples vs denser samples: what changes?
Adding samples further out with the same spacing: same FOV, finer pixels.
Putting samples closer together over the same range: larger FOV, same pixel size.
Both need more samples, and in the phase direction that means more TRs (more time).
7

What happens if every second phase-encode line is skipped (no correction)?
The spacing Δky doubles, so FOV_y halves. Anything outside the new, smaller FOV wraps into the image (aliasing).
The pixel size stays the same, because the outermost ky is unchanged.
8

Why does wraparound happen? (worked example)
Phase offset repeats every 360°, so a position outside the phase FOV ends up with exactly the same phase offsets as a position inside it. The scanner puts its signal there, on top of the real tissue.
Worked: FOV from −10 to +10 mm (20 mm wide). Signal at +12 mm appears at 12 − 20 = −8 mm.
Wraparound is mainly a phase-direction problem, because the readout direction is protected by filtering and oversampling.
9

What happens if outer k-space is cut off but the central spacing is kept?
The FOV stays the same (same spacing), but the image loses fine detail and looks blurred, because the outer points carry the fine detail. Cutting off abruptly also gives ringing (ripples) next to sharp edges.
This is the Week 4 demo of cutting k-space down from 512×512 to 8×8.
10

What is the point-spread function (PSF)?
The image you would get of a single tiny point with your sampling. It describes how the acquisition smears each point, so it belongs to the sampling, not to the object. Every image is the real object smeared by the PSF.
A broad main peak means blur, side ripples mean ringing, and extra copies mean aliasing. With few k-space points (as in MRSI) the PSF ripples, so signal leaks into neighbouring voxels.
11

How much phase does the largest phase-encode step put between neighbouring pixels?
About 180° (half a cycle), so neighbouring pixels end up with opposite phase offsets. That is the finest pattern the grid can show, which is why the largest step sets the resolution.
Not to be confused with the smallest step ΔGy·τ, which gives one full 360° cycle across the whole FOV.
12

What does keeping only k = 0 give, versus putting the same value at every k?
Only k = 0: a perfectly uniform image showing just the average brightness, because k = 0 is the whole slice added up with no phase offsets.
The same value at every k: a single bright point. A point in one domain is spread everywhere in the other.
13

What is the trade-off of apodization (tapering outer k-space, e.g. a Hamming filter)?
It turns the outer k-space points down gradually instead of cutting them off. This reduces ringing (smaller side ripples in the PSF) but makes the PSF broader, so the image is more blurred.
Less ringing, but also less detail.
14

How does voxel size affect SNR, e.g. halving slice thickness?
Signal comes from the protons in the voxel, so SNR is proportional to voxel volume. Halving the slice thickness halves the signal per voxel, so SNR roughly halves if nothing else changes.
The gain is finer detail through the slice and less mixing of tissues in one voxel. More slices may be needed to cover the same region.
15

How much averaging recovers a factor-2 loss of SNR?
4 times as many averages, so about 4 times the scan time, because SNR grows with the square root of the number of averages.
Averaging only reduces random noise. It cannot fix wraparound or create contrast that is not there.
16

Larger FOV with the same matrix: what happens?
More coverage, but each pixel is larger (coarser), so there is more signal per voxel. The number of phase encodes is the same, so the scan time is the same.
To keep the original pixel size you would also need a larger matrix.
17

Fewer phase encodes at the same FOV: what happens?
Fewer TRs, so a faster scan, but coarser pixels in the phase direction, because ky does not reach as far out.
The FOV is unchanged, because the spacing between steps is unchanged.
18

Why does cutting phase encodes save time but cutting readout points usually doesn't?
Each phase-encode step needs its own TR. All the readout points of a line are recorded during the same echo, within one TR.
So fewer readout points only shortens the readout, not the number of repetitions.
19

Scan-time formulas and worked example (TR = 1.7 s, 128 phase encodes)
2D: scan time = TR × number of phase encodes × number of averages.
3D: scan time = TR × N_y × N_z × number of averages.
Worked (Practical 2): 1.7 s × 128 = 218 s ≈ 3.6 min in 2D. As 3D with 128 partitions: 1.7 s × 128 × 128 ≈ 27,850 s ≈ 7.7 h.
20

How can many slices be acquired within one TR?
By interleaving. While slice 1 recovers during its TR, the scanner excites and reads slices 2, 3 and so on, then comes back to slice 1 for its next phase-encode step.
Extra slices are free until they no longer fit inside one TR. After that you need a longer TR or a second package. A longer TE makes each slice take longer, so fewer fit. (Practical 2 compared 30 and 120 slices.)
21

What causes slice cross-talk, and how is it reduced?
Real slice profiles have soft edges, so exciting one slice also partly tips spins in its neighbours. When a neighbour's turn comes it has less Mz, so it looks darker.
Reduced by: 1) gaps between slices, 2) exciting slices in interleaved order (odd then even, so neighbours have more time to recover), 3) pulses with sharper slice profiles.
22

How can phase-direction wraparound be reduced? (Practical 2)
1) Make the phase FOV larger, or oversample in the phase direction.
2) Swap the phase and readout directions, so the phase direction runs along the object's shorter side.
3) Put saturation bands over the tissue outside the FOV.
A bigger matrix with the same too-small FOV does not help, because the spacing, and so the FOV, is unchanged.
23

How does a saturation band remove signal?
1) A slice-selective pulse of about 90° tips Mz in the unwanted slab.
2) A spoiler gradient scrambles that Mxy so it gives no signal.
3) The imaging pulse comes right after, finds Mz ≈ 0 in the slab, and gets no signal from there.
Mz recovers with T1, so the band has to be repeated every TR.
24

Saturation bands: placement, dark margins and cost
Place the band over the tissue causing the problem (e.g. the tissue that wraps), but keep its soft edge away from the region you want, otherwise wanted tissue next to it turns dark.
Each band adds RF pulses, so more SAR, more load on the hardware and more preparation time.
25

Gradient heating vs RF hardware limits vs SAR
Three different limits.
Gradient heating: the large currents heat the gradient coils, limited by cooling and how often the gradients run.
RF hardware limits: the maximum power the amplifier, cables and coil can take.
SAR: the RF energy absorbed by the sample (W/kg), a safety limit.
More saturation or refocusing pulses raise both the RF load and SAR.
Alle 25 Karten kostenlos lernen

Mehr Karteikarten: Medizin und Gesundheit

Immunopharmacology134 KartenMedizin und Gesundheit · EnglischCoryMetabolism and Abdominal Anatomy94 KartenMedizin und Gesundheit · EnglischfaithClinical Microscopy (P2 Fecalysis)73 KartenMedizin und Gesundheit · EnglischMARIAHematology and Immune System (BIOE)57 KartenMedizin und Gesundheit · EnglischShaniaExcreción y Relaciones térmicas 47 KartenMedizin und Gesundheit · SpanischLuisaLecture 7 - Autoimmunity, Autoimmune Diseases37 KartenMedizin und Gesundheit · EnglischKim 3.8 nervous system30 KartenMedizin und Gesundheit · EnglischmischaColorații histologice și tipuri celulare24 KartenMedizin und Gesundheit · RumänischSourAcute Derm20 KartenMedizin und Gesundheit · EnglischMadiIrregular Affirmative/Negative15 KartenMedizin und Gesundheit · SpanischMatthewDrug - Pharmacodynamics10 KartenMedizin und Gesundheit · EnglischRyanTraumatic Brain Injury: Types and Effects189 KartenMedizin und Gesundheit · EnglischIrene

Alle Decks: Medizin und Gesundheit

Lerne diese Decks in der App

Kostenlos auf iPhone, Android, Mac und im Web. Füge jedes Deck mit einem Tipp zu deinem Konto hinzu.

Laden im App StoreJetzt bei Google Play
Alle 25 Karten kostenlos lernen